Description
Low Pressure: | Measurement: 0 to 50 psia (345 kPa) Pore Diameter: 360 to 3.6 µm |
High Pressure: | Measurement: Models 9500/9505 From atmospheric pressure to 33,000 psia (228 MPa) Models 9510/9520 From atmospheric pressure to 60,000 psia (414 MPa)
Pore Diameter: Models 9500/9505 6 to 0.0055 µm Models 9510/9520 6 to 0.003 µm
Transducer Accuracy: +/-0.10% of full scale (transducer manufacturer’s specifications)
Transducer Hysteresis: .05% of full scale for both the 33,000 and 60,000 transducer |
Penetrometers: | Capillary Stem Intrusion Volumes: 0.38, 1.1, 1.7, 3.1, and 3.9 cm3
Intrusion Accuracy: +/-1% of full scale intrusion volume
Sample Size: Maximum: a cylinder 2.5 cm in diameter by 2.5 cm long (1 in. diameter by 1 in. long) |
Equilibration Techniques: | By Time: 0 to 10,000 seconds By Rate: 0 to 1000.000 µL/g per second By Scanning: Continuous |